AeroTrak 9550 Specifications |
| Size Range | 0.5 to 25 μm |
| Particle Channel Sizes | 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 μm (additional channel sizes available) |
| Size Resolution | under 15% @ 0.5 μm (ISO 21501-4) |
| Counting E?ciency | 50% at 0.5 μm, 100% for particles > 0.75 μm (per ISO 21501-4 and JIS) |
| Concentration Limit | over 200,000 cubic feet @ 5 percent coincidence loss |
| Light Source | Laser diode (long life) |
| Zero Count | under 1 count / 5 min. - ISO 21501-4 & JIS B9921 |
| Flow Rate | 3.53CFM (100 Liters / minute) +/- 5 percent accuracy (ISO 215014 & JIS B9921 compliant) |
| Control of Flow Rate | Automatic, electronic closed loop (patented ?ow control) |
| Extension for Sampling Tube | 26.2 feet (8 meters) max. |
| Sample Modes | beep, automatic, manual; cumulative / di?erential; concentration or count |
| Sample Time | 1 sec. - 99 hrs |
| Sampling Freq. | continuous or 1 - 9999 cycles |
| Exhaust | HEPA ?lter - Internal |
| Vacuum | Internal pump |
| Communications Mode | Mod bus TCP over USB or Ethernet |
| Storage fo Data | 10K sample records include time, date, 6 channels for particles sample volume, ID, and ?ow; transferable via TrakPro Lite PC software or USB data downloading |
| Alarm | Built in - under 85dB @ 1m - adjustable |
| Status Indicator | Laser, ?ow, low battery |
| Output for Alarm | Dry contacts, closed when alarm is engaged |
| Alarm Limits | Programmable for all particle channels |
| Display | VGA 5.7-inch (14.5 cm) touch-screen display |
| Languages | English, German, French, Spanish, Japanese, Chinese (simpli?ed), Italian |
| Printer | Thermal printer Built-in (available w/ out printer) - in avaialable languages |
Environmental Sensor Interfaces | Support for TSI test probe - 966, 964, 962, 960 (air velocity / temperature / RH) & Temp. / RH probe |
| Storage Range | 32 - 122 deg F (0 - 50 deg. C), to 98 percent RH noncondensing |
Resistance to External Chemicals | hydrogen peroxide, chlorinated solution, Isopropyl alcohol |
Environmental Sensing Interface | Support for TSI relative humidity, temperature & air velocity probes |
| Software compatibility | TrakPro Lite Secure, TrakPro Lite & FMS 5 |
| Unit ID | IP address con?gurable |
| Security | 2 level password for locking out usage & con?g. |
| Location ID | Up to 999 Locations; 16 characters long |
| Reporting | Pass / Fail for FS209E, EU GMP and ISO 14644-1 |
| Calibration | NIST-traceable with TSI calibration calibration once per year recommended |
| Size (HxWxD) | 9.5 x 9.0 x 9.2 inches (24.1 x 23.0 x 23.2 centimeters) w/out isokinetic inlet |
| Weight | 14.2 pounds (6.44 kilograms) w/ 2 batteries |
| Power | universal 110 - 240 Volts AC |
| Battery | Removable / rechargeable Lithium Ion - 2 slots |
| Battery operation time | over 2 hrs continuous |
| Recharging | 4 hrs (idle) |
| Operating temp. | 41 - 95 deg F (5 - 35 deg C) 20 - 95 percent non-condensing |
| Standards | JIS B9921, CE, ISO 21501-4 |
| Warranty | 2 yrs, extended warranty available |